korallenwächterkalibrierungfeinanpassungicp

Fine Adjustment: Making the Korallenwächter Even More Precise

After initial calibration, the Korallenwächter can be further refined via fine adjustment — for example by comparing it against an ICP analysis.

What Is Fine Adjustment?

Fine adjustment (formerly: post-calibration) is not a replacement for initial calibration — it is an optional next step. It allows you to fine-tune the device for individual circumstances such as tube length or device positioning, and to align it with an external reference value if needed.

Typical use case: Your Korallenwächter shows KH 7.8 °dKH, your current ICP analysis shows 8.1 °dKH. With fine adjustment you can permanently store this difference.

Warning

Fine adjustment is intended for small corrections. If the deviation between the device and reference is very large (more than 0.5–1.0 °dKH for KH), this indicates an installation error or technical problem — please contact support in that case.

How It Works

In the Typus app, you enter two values:

  1. The current measured value from the Korallenwächter
  2. A reference value (e.g. from an ICP analysis or certified reference solution)

The app calculates a correction factor from these and applies it to all future measurements. You typically only need to do this once.

Good Reference Sources

Reference Suitability Note
ICP analysis Very good Sample must be taken simultaneously and from the same location
Certified reference solution Good Direct control without time delay
Another test kit Conditional Only if that kit is itself calibrated

When Is Fine Adjustment Worthwhile?

  • After initial operation, when you have a reliable reference value
  • After changing tubes or other installation changes
  • When you consistently observe a small, constant deviation from a trustworthy reference value

What to Expect After Fine Adjustment

Small residual deviations from the target value are normal. Perfect agreement is not a realistic goal — stable, consistent measurements are.

Tip

For a meaningful ICP comparison: take the sample for the ICP and simultaneously start a measurement with the Korallenwächter — at the same location in the tank. Deviations of 0.1 °dKH for KH or ±20 mg/l for calcium are normal and do not require adjustment.